Description
Nellcor N 600 Features
The Nellcor Pulse Oximeter can store up to 48 hours of 4- second trending data or 24 house f 2 second trending data, this can then be transferred to pc or printed.
- Simple, intuitive operation and easy-to-see display with dimmer and contrast adjustment
- Three display options: plethysmographic waveform, large numerics, and customizable real-time trend
- On-screen quick guide provides instant help to assist caregivers with monitor operation
- Convenient selection of adult or neonate alarm, and user-configurable power-on settings
- Fast-averaging mode (2 to 3 seconds) to more precisely track rapidly changing saturations
- On-screen viewing of 24 to 48 hours of SpO2 and pulse rate trends taken at 2-4 second intervals; printing capability
- Variable pitch beep tone enables clinicians to hear changes in SpO2
- Backup speaker with escalating alarm tones
- Interfaces with certain multi-parameter monitors enabling real-time patient information and true physiologic waveforms to be displayed
- Works with the Oxinet III remote respiratory monitoring system
- 7-hour battery life with battery gauge for convenient portability
Nellcor N 600 Specifications
Dimensions
- Height: 4.1 in (104mm)
- Width: 11.9 in (302mm)
- Depth: 10.1 in (257mm)
- Weight: 13 lbs (6kg)
Display Type
- vacuum fluorescent & LEDs. Display are: 140.6mm x 35mm. Resolution: 256 x 64 pixels. Pixel size: 0.4 x 0.4 mm
Controls and Indicators
- Front panel: on/standby; freeze; menu; pulse tone volume, up and down; alarm off; alarm off indicator; battery charging indicator; function keys (4), software configured. Rear panel: mains AC power ON/OFF control.
Connections
- Front panel: connector for mainstream CO2 sensor cable; connector for SpO2 cable. Rear panel: power cord receptacle; fuse block; RS-232 input/ output port; C-lock ECG synchronization input (3/32” mono phone jack).
Capnograph Display Range
- 0-99 mmHg. Breath rate: 0-150 breaths/min. Accuracy: ±2 mmHg for 0-40 mmHg, ±5% of reading (increasing linearity 0.8% for every 10 mmHg) for 41-99 mmHg. Response time: less then 60 ms. Alarms: apnea; sensor dis connect; high/low ETCO2; high/low breath rate; high/low inspired CO2.
Pulse Oximeter Display Ranges
- 0-100% SpO2, 20-250 beats/min pulse rate. Accuracy: Adults: 70%-100% ±2 digits, 0-69% unspecified; Neonates: 70%-95% ±2%, 0-69% unspecified. Wavelengths: 660 nm (red, nominal); 920 mm (infrared, nominal). Heat dissipation: non-heated: less than 50 mW total heat dissipation by LEDs (typically less than 1° C temperature rise). Alarms: pulse search; high/low SpO2; high/low pulse rate; SpO2 sensor disconnect.
Co2 Sensor Type
- mainstream. Size: <12.5 mm3 (0.5 in3). Weight: less than 10 grams. Cable length: 3 meters (9.84 feet). Calibration: annual accuracy check; no user calibration required. Temperature: 42°C nominal
Electrical Power requirements
- 00/120V ~, 50/60 Hz, 500 mA; 220/240V ~, 50/60 Hz, 250 mA. Digital/Ana log input voltage: ±30 VDC, maximum. Leakage current: 10 μA maximum (for 110V; 20 μA maximum for 220V) from patient connections to ground or AC mains; 100 μA maximum from AC mains to chassis ground.
Battery Type
- Sealed lead acid. Operating time (fully charged): 90 min. nominal, 180 min.without graphics. Recharge period: 12 hrs typical. Charger type: internal, current limited, constant voltage.Battery fuse: 1XT 2.5 A, 250V, 3 AG.
AC Fuses
- 100/120 V ~ 1 XT 1.0 A, 250V, 3AG; 220/240 V~ 2 X T 0.5 A, 5×20 mm
Reviews